Content
Handbook H
BENFURACARB
OUTLINE OF METHOD
The sample is dissolved in acetonitrile. The components
are separated by high performance liquid chromatography on a reversed
phase column (C18) with acetonitrile + water (75 + 25 v/v) as mobile
phase and UV detection at 280 nm. Benfuracarb is determined by peak
area measurement using n-nonaphenone as internal standard.
SCOPE:
The method
is usable for TC and EC-formulations