Content
Handbook G
MECOPROP-P
OUTLINE OF METHOD
Mecoprop-P is dissolved in acetonitrile-water and chromatographed
on a HPLC reversed phase system with acetonitrilewater acetic acid
as mobile phase using UV-detection at 220 nm and external standardisation.
The enantiomeric ratio is determined after esterification by chromatography
on a chiral phase using n-heptane-n-pentanol as mobile phase and UV-detection
at 229 nm.
SCOPE:
The method
is usable for TC, WP and SL formulations.
last update
01.12.2010